Titra
Lab
®
TitraLab 870 Reference Manual
Page 184
QC sample
Quality control (QC) samples are used as a means of studying the
variation within and between batches of a particular analysis. A typical
QC sample will be stable, homogenous, typical in composition to the
types of sample normally examined. The concentration of a QC sam-
ple is known accurately and its composition is as close as possible to
one of the samples to be analysed.
Quality control samples can be used for a measurement or a titration
method.
QC sample (Yes/
No)
Select Yes for QC sample if you wish to perform measurements on
QC samples.
The QC sample ID, the periodicity of QC samples and the minimum
and maximum acceptance limits for the control test are entered in the
QC data menu.
If the QC test fails, the method can be locked, so that it is impossible
to run the method while the QC sample results lie outside the preset
limits.
Enter in:
Edit method menu
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...