Page 54
Auxiliary input
The auxiliary input socket can be connected to an external device unit
used to send an analysis start command to the Titration Manager.
The analysis is a sequence of methods with manual change of the
titration beakers (Working mode = Sequence,
).
The external device unit is to be connected to the red and black
IN
banana sockets of the Titration Manager. The red banana socket
receives the TTL 0 ± 5 V auxiliary signal and the black banana socket
is connected to the instrument electrical zero.
Proceed as follows to trigger a sequence of methods by an auxiliary
signal input:
•
In the Configuration menu, select
Controlled by TTL IN = Yes
.
•
Connect the auxiliary control unit to red and black
IN
banana
sockets of the Titration Manager.
•
Run the sequence. The Titration Manager displays a waiting for
auxiliary signal message.
The sequence is started as soon as the
auxiliary signal is received.
Spécifications of the auxiliary input signal
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...