Titra
Lab
®
TitraLab 870 Reference Manual
Page 75
Cell constant
(definition)
The TIM870 calculates and displays the
κ
conductivity based on a
G
measured conductance and the
K
cell constant of the conductivity cell
used.
κ
(in S
.
cm
-1
) = K x G (in S)
The
K
constant (expressed in cm
-1
) is a characteristics of the
conductivity cell depending on the cell geometry.
To measure conductivities, you must know the cell constant value.
With the TIM870, you can directly enter
K
in the Edit electrode menu
see "Cell constant (parameter)", page 74
) or
determine
K
by calibrating the conductivity cell
see "Electrode calibration (conductivity)", page 108
Cell grounding
Defines the grounding of the measuring cell. Select one of the
following options:
Reference
Grounding is ensured by a reference electrode -
general use
.
Metal
Grounding is ensured by a metal electrode connected to the
GND
socket on the Titration Manager. Use this option in case of high
resistive solutions in order to avoid measuring background noise at
the electrodes.
Others
Grounding is not ensured by the reference electrode and is defined
outside the method.
Enter in:
Edit method menu
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...