TitraLab 870 Reference Manual
Page 96
Direct ISE
measurement
method - notes
The accuracy of the measurements using a Direct ISE method
depends on the following elements:
•
The concentrations of the standards used for a 2-point calibration
must lie on either side of the samples to be measured.
•
For calibration using more than 2 standards, one of the standard
concentration must lie in the non-linear response zone of the ISE
electrode.
•
If you want to measure low concentrations (values situated in the
non-linear response zone), run a 2 or 3-point calibration in the non-
linear response zone of the electrode.
•
It is recommended to measure sample concentrations above the
C
0
limit.
•
A high value found for C
0
may undergo false measurements (check
your standards and electrode).
•
A similar ionic strength must be found in both standards and
samples (add a supporting electrolyte in the standards and
samples).
•
The samples must not contain a significant amount of interfering
ions.
•
Use the same temperature for your standards and samples
(thermostate the solutions).
How to edit a Direct ISE measurement method?
see "Direct ISE measurement method - programmation", page 96
.
How to run a Direct ISE measurement method?
Refer to "Running a method", page 216
.
What is a Direct ISE measurement method?
see "Direct ISE measurement method - definition", page 95
Direct ISE
measurement
method -
programmation
Proceed as follows to edit a Direct ISE measurement method:
1.
From the Main window, press
4
then
2
Edit method.
2.
For
Mode
, select
Measurement
.
3.
For
Measurement
, select
Direct ISE
.
4.
Define the other parameters of this measurement method.
What is a Direct ISE measurement method?
see "Direct ISE measurement method - definition", page 95
How to run a Direct ISE measurement method?
Refer to "Running a method", page 216
.
Direct
measurements
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...