- 42 -
indication is close to zero. If results vary less than the declared AUTOZERO range e.g. one division,
balance zeroes automatically, marker of stable measurement result –
,
and precise zero marker –
are displayed.
If AUTOZERO function is enabled, then each weighing process starts from precise zero point. There
are, however, some cases when this function can be a disturbing factor for the measuring process;
e.g. very slow placing of a load on the weighing pan (load adding). Here, zero indication correction
can also correct actual indication of loaded mass.
Last digit display
Function enables/disables displaying the last digit of decimal place for a weighing result.
Balance ambient conditions
Parameter relating to ambient and environmental conditions in which the balance operates. There
are two options: STABLE and UNSTABLE. Selecting STABLE mode makes the balance work much
faster, i.e. weighing takes much less time than for UNSBABLE mode. If the ambient conditions are
unstable it is recommended to use UNSTABLE mode. By default, the parameter is set to STABLE
option.
11.2.6. AUTOTARE
Autotare is used for quick determination of net weight for loads with different tare values, wherein
they are measured one after another.
When the function is active (<AUTOTARE> parameter set to <YES> option), the operating process
takes the following steps:
•
Make sure that the weighing pan is empty and press button responsible for zeroing.
•
Put product packaging on a weighing pan.
•
After measurement stabilization,
automatic taring
of the packaging mass proceeds (
Net
marker appears in the upper part of the display).
•
Put product that is to be packed into the packaging.
•
The display shows net weight of the product.
•
Take off the product together with the packaging.
•
The balance cancels tare value (packaging weight recorded in balance storage during the first
step of the operating process) after the gross weight value (set in
<AUTO THRES>
parameter)
has been exceeded.
•
Put packaging of the next product on a weighing pan, automatic taring of the packaging weight
proceeds after measurement stabilization (
Net
marker appears in the top section of the
display);
Summary of Contents for AS R2.PLUS
Page 6: ...6 1 GENERAL INFORMATION 1 1 DIMENSIONS AS PLUS series PS 200 2000 PS 1000 PS 2100 PS 10100...
Page 97: ...97...
Page 98: ...98...