- 69 -
•
Enter (V2) control volume. The computer software proceeds to the next step, i.e. entering the
third control volume (V3) for a tested pipette.
•
Enter (V3) control volume.
•
Upon entering the data the computer software proceeds to pipettes calibration.
•
Follow the displayed description and complete the procedure.
•
By means of a pipette, dose the first portion of water, wait until the measurement is stable and
press ENTER to confirm.
•
The computer software records the measurement. If P2.3.7
AUTOMATING TARING
parameter
is set to value <YES> then the indication is automatically tared. If P2.3.7
AUTOMATIC
TARING
parameter is set to value <NO> than the user must press TARE button in order to tare
the indication before dosing another portion of water.
•
Upon completion of series of measurements, summary is displayed.
•
Press ENTER to continue measuring process, press Esc. to stop it. Upon pressing Esc. button
the computer software returns to the main window.
•
Upon pressing ENTER, while displaying the summary result for the last adjustable pipette
volume, a report is generated. The report is printed on a printer connected to the balance
(ambient conditions visible on a report are those that user entered at the beginning of the
calibration procedure). For fixed volume pipette the situation is similar if the user presses
ENTER while displaying summary for volume V1.
•
The computer software returns to the main window.
•
Now the user may either start new procedure for the same pipette or enter new data for a
different pipette.
An exemplary report: adjustable volume pipette, 3 volumes tested: