- 82 -
•
unstable – on pressing
button, display status is sent immediately to a peripheral device (on
a printout such status is marked with <?> symbol located in front of the weighing result). This
option is only available for non-verified balances.
14.1. TRANSFERRED DATA FORMAT
The measurement result can be transferred from a balance to an external device by pressing
button, located on the balance, or by sending a control command from a computer.
14.2. FORMAT OF DATA SENT ON PRESSING PRINT BUTTON
CAUTION!
Unstable measurement printout is disabled for a verified balance.
Printout format for unverified balances:
1
2
3
4 - 12
13
14 - 16
17
18
s
tabi
lit
y
m
ar
k
er
s
pac
e
c
har
ac
ter
ma
s
s
s
pac
e
uni
t
CR
LF
Stability marker
[space] if stable
[?] if unstable
[^] if high limit is out of range
[v] if low limit is out of range
Character
[space] for positive values
[-] for negative values
Mass
9 characters with right justification
Unit
3 characters with left justification
Printout format for verified balances:
1
2
3
4 - 14
15
16 - 18
19
20
s
tabi
lit
y
m
ar
k
er
s
pac
e
c
har
ac
ter
ma
s
s
s
pac
e
uni
t
CR
LF
Stability marker
[space] if stable
[?] if unstable
[^] if high limit is out of range
[v] if low limit is out of range
Character
[space] for positive values
[-] for negative values
Mass
11 characters with right justification (contains decimal marks)
Unit
3 characters with left justification