- 16 -
5. START-UP
When plugged to mains, the balance displays program name and number, next it proceeds to the
weighing mode.
5.1. TEMPERATURE STABILIZATION PERIOD
Before start of measuring processes, it is necessary to wait until the balance reaches thermal
stabilisation.
For balances that were stored in much lower temperatures before plugging to mains (e.g. during
winter period), thermal stabilisation period shall take at least 4 hours for PS balances, and 8 hours for
AS balances. During the thermal stabilization, the indications on the display panel can change.
It is recommended that ambient temperature changes at place of use were insignificant (slow to
change).
5.2. AMBIENT CONDITIONS STATE INDICATION
The function is intended to inform on unstable ambient conditions for a balance, it is enabled only for
AS R series balances.
The function controls dynamic temperature changes occurring in the balance during its operation. If
the variation is greater than set limit values (temperature changes speed), then a blinking
thermometer pictogram is displayed on the screen.
The blinking thermometer pictogram means that temperature inside the balance is not stable, this
may result in inaccurate mass measurement. For such a case it is recommended to wait until the
temperature stabilizes or to perform balance adjustment (blanking of the blinking thermometer
pictogram).
5.3. USER MENU
Menu is divided into 9 basic function groups. Each group has an individual name starting with
a capital letter,
P
.
P1 ADJUSTMENT
P1.1
INT. CALIB.
|
[internal adjustment]
P1.2
EXT. CALIB.
|
[external adjustment]
P1.3
USER CALIBRATION
|
[user adjustment]
P1.4
CALIBRATION TEST
|
[adjustment test]
P1.5
AUTO. CALIB.
|
[automatic adjustment]
P1.6
AUTO. CALIB. C.
|
[time of automatic adjustment]
P2 WORKING MODES
P2.1
ACCESSIBILITY
|
[settings for the accessibility of individual modes
while working with the balance]
P2.2
WEIGHING
|
[setting for the function weighing]
P2.3
COUNTING PCS
|
[settings for the function counting pieces]
P2.4
CHECKWEIGHING
|
[settings for the function checkweighing]
Summary of Contents for AS R2.PLUS
Page 6: ...6 1 GENERAL INFORMATION 1 1 DIMENSIONS AS PLUS series PS 200 2000 PS 1000 PS 2100 PS 10100...
Page 97: ...97...
Page 98: ...98...