- 51 -
<Ok>: mass contained within thresholds
<Max>: mass higher than the value of the high threshold
Since LCD display capacity is limited, the threshold markers do not reflect threshold setups and
target weight value precisely. They provide rough information and serve as user aid in course of the
balance operation.
11.5. DOSING
Dosing mode comprises process of sample weighing, wherein the said sample weighing is performed
until target mass is reached. The target mass is defined along with dosing tolerance value. The
tolerance value is set as a percentage of the target mass.
An example:
Target weight = 100.000g
Tolerance = 2,5% (2,5% of 100g, which amounts to 2,5g)
i.e.: the software accepts correctly dosed value within the following thresholds: from 97,500g to
102,500g.
F Shortcut Keys
You can declare which particular weighing function is to be launched using F1, F2, F3 or F4 keys.
For procedure informing you how to declare the particular function, read
F Shortcut Keys
section.
11.5.1. Setting Target Mass: Entering Mass Value
Procedure:
•
Press F button to which function <
ENTER SAMPLE
> is assigned, the editing window is
displayed. The target weight value must be given in a current unit.
•
Use arrow buttons to enter the target weight, press
button to confirm.
•
The software automatically sets tolerance of target mass dosing. Use arrow buttons to set the
tolerance value, press
button to confirm.
•
The balance software automatically enters <
DOSING
> mode and displays the target mass
value with a minus sign, and in the bottom line, reference mass value – target mass (if such
option has been selected for <
INFORMATION
> function).
Summary of Contents for AS R2.PLUS
Page 6: ...6 1 GENERAL INFORMATION 1 1 DIMENSIONS AS PLUS series PS 200 2000 PS 1000 PS 2100 PS 10100...
Page 97: ...97...
Page 98: ...98...