- 68 -
Additional settings list:
•
VOLUME DETERMINATION
– allows to predefine number of tested volumes for a particular
pipette. For pipettes with fixed volume please set parameter <1>, for pipettes with adjustable
volume set parameter <2> and <3>.
•
MEASUR. NO
– allows to predefine number of measurements for each tested volume. Number
of measurements ranges from 6 to 20.
•
AUTOMATIC TARING
- allows to turn on function of automatic taring of dosed portion of water
after confirmation of measurement (value set to <YES>)
Remember to select correct values of the above options before pipettes calibration procedure is
performed. The settings should reflect expectations and needs being a result of working environment.
F Shortcut Keys
You can declare which particular weighing function is to be launched using F1, F2, F3 or F4 keys.
For procedure informing you how to declare the particular function, read
F Shortcut Keys
section.
11.14.2. Means of Operation
•
Enter <PIPETTES CALIBRATION.> mode.
•
Start the process.
•
Follow the prompts to set the correct options.
•
Enter the ambient temperature, next press ENTER for confirmation. The computer software
proceeds to the next step, i.e. ambient humidity settings.
•
Set the ambient humidity, next press ENTER for confirmation. The computer software proceeds
to the next step, i.e. air pressure settings.
•
Set the air pressure, next press ENTER for confirmation. The computer software proceeds to
the next step, i.e. entering the first control volume (V1) for a tested pipette.
•
•
Enter (V1) control volume. For fixed volume pipettes (P2.13.5 VOLUME DETERMINATION
parameter set to value <1>) this is the one and only value to be entered. The computer
software proceeds to the next step, i.e. entering the second control volume (V2) for a tested
pipette.