- 75 -
13. PERIPHERAL DEVICES
PERIPHERAL DEVICES menu is comprised within Parameters menu. It is accessed by pressing
key. The menu features list of devices that can cooperate with the balance.
13.1. COMPUTER
<COMPUTER> submenu allows the user to select a port for connection of a computer running
software which enables:
•
communication with a balance,
•
switching the continuous transmission on and off, and
•
cooperation of a balance with E2R software of Radwag production.
13.1.1. Computer Connection Port
Procedure:
-
press
button
-
enter <P4 DEVICES> menu
-
enter menu group: <P4.1 COMPUTER>
-
select port to which a computer is to be connected,
COM 1
or
COM 2
– RS 232 port, for connecting a computer
USB 2
type B – USB port, for connecting a computer
USB
FREE LINK
- USB port, type B, for connecting a computer. Tool for entering data, it
functions as a keyboard. Upon proper modification of non-standard printout and sending
proper command from the computer, or upon pressing ENTER key located on the
operation panel, the data from non-standard printout is entered directly to computer
programs, e.g. Excel, Word, Notepad etc.
Select <USB FREE LINK> port and configure its settings in order to enable correct
cooperation of the balance with computer programs:
SPREADSHEET – value ‘YES --.—‘ – cooperation with Excel type program, transferred
decimal values are separated by dot (.)
– value ‘YES --,--‘ – cooperation with Excel type program, transferred
decimal values are separated by coma (,)
Summary of Contents for AS R2.PLUS
Page 6: ...6 1 GENERAL INFORMATION 1 1 DIMENSIONS AS PLUS series PS 200 2000 PS 1000 PS 2100 PS 10100...
Page 97: ...97...
Page 98: ...98...