- 90 -
working mode.
Format: UI CR LF
Response options:
UI_”x
1
,x
2
, … x
n
”_OK CR LF
– command carried out, returns accessible units
UI_I CR LF
– command understood but not accessible at this
moment
x
– unit symbols, separated by means of comas
x
→ g, mg, ct, lb, oz, ozt, dwt, tlh, tls, tlt, tlc, mom, gr, ti, N, baht, tola, u1, u2
An example:
Command:
UI CR LF
– return available units
Response:
UI_”g, mg, ct”_OK CR LF
– response: available units
Set current unit
Command overview: Command sets current unit for a particular device.
Format: US_x CR LF
Response options:
US_ x_OK CR LF
– command carried out, response: current unit
US_E CR LF
– error in-course of command execution, no parameter or incorrect
format
US_I CR LF
– command understood but not accessible at this moment
x
– parameter, units symbols: g, mg, ct, lb, oz, ozt, dwt, tlh, tls, tlt, tlc, mom, gr, ti, N, baht, tola, msg,
u1, u2, next
CAUTION! if x=next, the command swaps to another avaialable unit on the list (it simulates „
” button
pressing).
An example:
Command:
US_mg CR LF
– set „mg” unit
Response:
US_mg_OK CR LF
– „mg” set as a current unit
Give current unit
Command overview: command returns current unit.
Format: UG CR LF
Response options:
UG_x_OK CR LF
– command carried out, response: current unit
UG_I CR LF
– command understood but not accessible at this moment
x
– parameter, unit symbol
An example:
Command:
UG CR LF
– return current unit
Response:
UG_ct_OK CR LF
– currently set unit is „ct”
Activate sound signal
Command overview: command activates BEEP sound signal for a specified amount of time
Format: BP_CZAS CR LF
Response options:
BP_OK CR LF
– command carried out, BEEP sound signal activated
BP_E” CR LF
– no parameter or incorrect format
BP_I CR LF
– command understood but not accessible at this moment
TIME
– parameter, decimal value specifying how long shall the sound last, parameter given in [ms].
Recommended range <50 ÷ 5000>.
If value greater than the permissible high limit is given, than BEEP sound is operated for the
maximum permissible amount of time.
An example:
Command:
BP_350 CR LF
– activate BEEP for 350 ms
Response:
BP_OK CR LF
– BEEP activated
CAUTION!
BEEP sound activated via BP command is inhibited if in-course of its activation the sound is activated by
means of other device: keypad, touch panel, proximity sensors.