- 24 -
04 Adjustment test
mass comparison of internal adjustment weight with its value saved in balance
memory.
05 Internal weight correction
The function enables correcting the value of internal adjustment weight. The function is
disabled in verified balances.
06 Automatic adjustment
Determination of factors triggering start of automatic internal adjustment:
0
no
– none of factors will cause start of adjustment
1
time
– adjustment triggered by time interval set in point 07
2
temperature – adjustment triggered by change of temperature
3
both
– adjustment triggered by time and temperature
07 Time of automatic adjustment
Determination of time interval, after which automatic adjustment process is initiated.
Return to weighing mode
Changes introduced in balance memory will be saved on returning to
weighing with procedure of saving changes. Press ESC key for a few times
until the display shows a question: Save? As displayed, select one of
available options:
ENTER – save changes and go back to menu;
ESC
– abandon changes and go back to menu.
(see Fig. 9. Return to weighing mode. point. 5.1.2. Return to weighing
mode).
7.2. Adjustment test
Adjustment test is a comparison of internal adjustment weight with its value stored in
balance’s memory. This process is carried out automatically and its result is shown on the
display.
Fig. 13. Adjustment test
Cal. – the value of internal adjustment weight
Curr. – weighing result of the internal adjustment weight
Dev. – deviation calculated between the both values