- 67 -
Fig. 54. Statistics – display content including a series of measurements
1. Mass currently placed on the weighing pan
2. Number of measurement in a measuring series
3. Sum of all completed measurements in a series
4. Average mass of all completed measurements in a series
5. Minimum saved mass in a carried out measuring series
6. Maximum saved mass in a carried out measuring series
7. The difference between the minimum and maximum saved mass in a measuring
series
8. Precisely calculated standard deviation value
9. Variance factor value
10. Measuring unit [g]
11. Working mode name
Statistics mode features a special option enabling quick calculating of statistics from a
measuring series with a possibility of generating reports from the completed calculations.
The user can declare the content of a header and footer in a printout of statistical report.
The header <PRINTOUT 1>, and the footer <PRINTOUT 2> are settable in non-standard
printouts.
In order to correctly use the option, the user has to create templates 1 and 2 using the non-
standard printouts (
see point
19.2
). As the templates are ready, the option can be enabled
in mode settings.
Means of operation:
−
Press
key to print the header (PRINTOUT 1)
−
Carry out a measuring series (set all options of the GLP menu to NO), and accept
each measurement result by pressing <ENTER> key
−
On completing a measuring series press F key to print a statistics from the series
−
In order to continue the measuring series accept it by pressing <ENTER> key or
−
Press
key to print a footer (PRINTOUT 2)