- 74 -
17.2.1.
Inserting text into strings
Variables available in all working modes and having the same value
%%
Printout of a single character “%”
%N
Current net mass in basic measuring unit
%d
Current date
%t
Current time
%i
Balance factory no.
%R
Software no.
%P
Project no.
%U
User no.
%F
Name of an active working mode
%C
Date and time of last completed adjustment process
%K
Type of last completed adjustment process
%I
Deviation in last completed adjustment process
%1
Code 1
%2
Code 2
%3
Code 3
%4
Code 4
%5
Code 5
%6
Code 6
Variables dependent on a currently enabled working mode
Variable Description
Working mode in which the variable is
active
%W
Mass of a single part PARTS COUNTING
%H
HI high threshold
CHECKWEIGHING
%L
LO Low threshold
%Z
Target mass
DOSING
%B
Reference mass
PERCENT SETUP
%A
Filter
ANIMAL WEIGHING
%b
Threshold (limit)
%i
Liquid
DENSITY DETERMINATION
%p
Procedure
%c
Temperature
%a
Liquid density
%v
Sinker volume