- 87 -
22. CONNECING SCHEME OF EXTERNAL BUTTONS
A balance AS/X series enables connecting external buttons for tarring or printing
measurement results without the need to touch balance’s keyboard with operator’s hands.
The buttons are connected using a dedicated splitter connected to balance’s RS 232
interface.
Additionally the splitter enables connecting a printer or a computer.
The splitter and buttons are additional equipment (optional) of a balance AS/X series.
23. LIST OF COMMANDS COMPUTER - BALANCE
Function
INTERFACE RESET
Command
R CR LF (zeroing currently carried out commands, restoring factory default
settings)
Function
SEND ALL IMPLEMENTED COMMANDS
Command
PC CR LF (sends information on all commands implemented in the
balance’s software)
Function
SEND MEASUREMENT RESULT IN BASIC MEASURING UNIT
Command
S CR LF (sends result in basic measuring unit on stabilization of indication)
Function
IMMEDIATELY SEND MEASUREMENT RESULT IN BASIC MEASURING
UNIT
Command
SI CR LF
Function
SEND MEASUREMENT RESULT IN CURRENT MEASURING UNIT
Command
SU CR LF sends result in current measuring unit on stabilization of
indication)
Function
IMMEDIATELY SEND MEASUREMENT RESULT IN CURRENT
MEASURING UNIT
Command
SUI CR LF
Function
ZERO BALANCE
Command
Z CR LF (zeroing of indication on stabilization)
Function
IMMEDIATELY ZERO BALANCE
Command
ZI CR LF
Function
TARE WHEN STABLE
Command
T CR LF
Function
IMMEDIATELY TARE BALANCE
Command
TI CR LF