- 88 -
Function
SWITCH OFF CONTINUOUS TRANSMISSION IN BASIC MEASURING
UNIT
Command
C0 CR LF
Function
SWITCH ON CONTINUOUS TRANSMISSION IN BASIC MEASURING
UNIT
Command
C1 CR LF
Function
SWITCH OFF CONTINUOUS TRANSMISSION IN CURRENT
MEASURING UNIT
Command
CU0 CR LF
Function
SWITCH ON CONTINUOUS TRANSMISSION IN CURRENT MEASURING
UNIT
Command
CU1 CR LF
Function
GIVE FACTORY NUMBER
Command
NB CR LF
Function
GIVE MEASURING RANGE
Command
FS CR LF
Function
GIVE SOFTWARE REVISION
Command
RV CR LF
Function
GIVE OR CHANGE DATE IN A BALANCE
Command
PD CR LF (causes sending data on set date or changes the date)
Function
GIVE OR CHANGE TIME IN A BALANCE
Command
PD CR LF (causes sending data on set time or changes the time)
Function
GIVE ENABLED WORKING MODE
Command
PM CR LF
Function
SEND SETUP
Command
PS CR LF (causes sending complete data on balance setup – printout of
parameters)
Function
“BEEP” SOUND
Command
B CR LF (causes immediate activating of beep sound in a balance)
Function
GIVE LAST CODE ERROR
Command
ER CR LF (causes sending code of last saved error in a balance)
Function
SEND TEXT STRING
Command
DS CR LF (causes previewing a sequence of characters)
Function
ERASE TEXT STRING
Command
CS CR LF (causes erasing a text string and restoring default display status)