- 116 -
Exemplary settings for connected ambient conditions module:
Baud rate – 115200 bit/s
Address - 15
26. AMBIENT CONDITIONS
Parameters group which has been designed to enable you to turn on readout of ambient conditions
recorded by THB 3/5 ambient conditions module, and to specify tolerance temperature and humidity
values and a change rate for the values per hour. Entered values are referred to indicted values.
Next on the basis of values comparison respective pictograms are displayed informing you whether
the sensor-read values are comprised within permissible limits or not.
Particular sensors settings:
•
Ambient conditions recording interval [min]: parameter enabling you to specify record frequency
of sensor-registered data, and to determine how often the displayed pictograms (providing
information on ambient conditions state) are to be refreshed,
•
Internal sensor 1: enter this parameter to specify tolerance temperature values and balance
temperature change rate,
Min temperature
– min balance temperature, for lower
temperature values the thermometer pictogram is red
Max temperature
– max balance temperature, for higher
temperature values the thermometer pictogram is red
Temperature
∆
t/h
– maximum balance temperature change
rate, for higher change rate values the balance displays blinking
red thermometer pictogram
In order to enable/disable parameter visibility go to service menu.
•
Ambient conditions module: parameters group which has been designed to enable you to turn
on readout of data recorded by a balance-connected ambient conditions module, and to specify
tolerance values for this module.