- 17 -
Remove a weighing pan, a draft
shield, a bottom insert. Clean the
components when detached. With
this your balance mechanism is
protected against accidental
damage.
Thus prepared draft shield and panes can be properly cleaned. All the operations should be done
carefully. Pay special CAUTION to the spot where the weighing pan was installed: dirt and other
small elements might enter the balance construction through this opening, which might negatively
influence the balance parameters.
3.5. POWERING THE DEVICE
Balance can be connected to the mains only with a power adapter that comes standard with the
particular model. Nominal power supply of the power adapter (specified on the power adapter data
plate) should be compatible to the power from the mains.
Plug the balance to the mains – connect the power adapter to the socket, next connect its connector
to port located at the back of the balance housing.
Test of the display unit takes place right after connecting the balance to the power, all the elements
and pictograms are backlit for a short time. Next, the name and the program number appears, the
indication gets to ZERO (displayed reading unit depends on the balance). During the balance start,
the test of an internal mass adjustment mechanism occurs (single location and elevation of the
internal mass adjustment).
If the indication is different than zero, please press
button.
CAUTION!
If the balance is “verified”, automatic adjustment occurs right after switching the balance on.
3.6. TEMPERATURE STABILISATION TIME.
Before start of measuring processes, it is necessary to wait until the balance reaches thermal
stabilisation.
For balances that were stored in much lower temperatures before plugging to mains (e.g. during
winter period), thermal stabilisation period shall take at least 4 hours for PS and WLC balances, and
8 hours for AS and APP balances. During the thermal stabilization, the indications on a display panel
can change. It is recommended that ambient temperature changes at place of use were insignificant
(slow to change).
3.7. CONNECTING ADDITIONAL HARDWARE
Use only accessories and peripheral equipment recommended by the manufacturer. The balance
must be disconnected from the mains before connecting or disconnecting any peripherals (printer,
PC computer, computer keyboard). On connecting the peripherals, plug the balance to the mains.