- 65 -
•
Enter the requested value and confirm by pressing
key, <Parts counting> mode is run
with automatically specified part mass.
CAUTION!
If you enter weight of value that is lower than 0.1 reading unit, then
<Value too low>
message is displayed.
13.4. SETTING REFERENCE MASS BY DETERMINING PART MASS
Procedure
:
•
Put a container on a weighing pan, tare its mass,
•
Press <
Determine part mass> button,
<Reference quantity
> editing window
featuring an on-screen keyboard is displayed,
•
Enter the requested value and confirm by pressing
key,
•
Message on number of parts to be loaded is displayed:
<Number of parts to be loaded: 18>
,
•
Put specified number of pieces on the weighing pan, when the result is stable (
stability
pictogram displayed) press
key to confirm,
Summary of Contents for AS X2 Series
Page 7: ...7 1 GENERAL INFORMATION 1 1 DIMENSIONS AS X2 series PS X2 series...
Page 8: ...8 APP X2 series WLC X2 series...
Page 96: ...96 Status Completed Signature...
Page 132: ...132...