19.2. ANIMAL WEIGHING – QUICK ACCESS KEYS
Each working mode features automatically displayed set of default keys. The set can be modified by
assigning out-of-list quick access keys to on-screen keys. Such operation requires particular
permissions level.
Read section 7.6.1 for list of quick access keys.
Procedure for animal weighing
•
Set the function parameters (averaging time, autostart, auto thres).
•
Load the object that is to be weighed on a weighing pan, press <
> key; for automatic
operation, mass determination process starts when the pre-set auto threshold mass is
exceeded.
•
On completed mass determination process, „freezed” weighing result is displayed, the mass is
automatically printed using the printer.
The weighing result can be printed repeatedly, press <
> key.
•
The next measurement can be carried out when <
> key is pressed:
for manual operation press <
> key,
for automatic operation, remove weighed object form the weighing pan and load the
new object.