- 98 -
•
The program automatically reads the file and the data is imported to the balance. Respective
prompts are displayed for confirmation of sucsessfully completed operation.
DELETE A RECORD
Procedure:
•
Press the record and hold it.
•
A message: <Confirm to delete> is displayed.
•
Press <
> to confirm, the record is deleted from the list.
DELETE A DATABASE
Procedure:
•
Enter selected database and press
(Delete everything) pictogram located in the top bar.
•
A message: <Confirm to delete all records> is displayed.
•
Press <
> to confirm, the database is deleted.
FORMULATIONS REPORTS, WEIGHING REPORTS and ALIBII cannot be imported. ALIBII
database cannot be deleted. The databases content can only be exported and saved to a flash drive.
Names of files with exported data consist of serial number and proper extension (see table below).
Database
File name and extension
Formulations Reports
123456.for
Weighing Reports
123456.wei
Alibii
123456.ali
Files can be read using ALIBI Reader by RADWAG that can be downloaded from the website:
www.radwag.pl.
AMBIENT CONDITIONS database serves informative purposes only. Using records preview option
you can check ambient conditions and observe how they changed over time. The data saved to
records can be printed by pressing
pictogram located in the top bar. Records’ names contain
date and time of saving the record’s data to balance memory.
23.2. PRODUCTS
Product database stores names of all products that can be weighed, counted, controlled.
Procedure:
•
Enter <
Databases> submenu, press <Products> key.
•
Press <
Add> key for adding a new product.
•
For operating an already introduced product press the product field.
List of parameters defined for a product:
1.
Name
2.
Code
[product code]
3.
EAN
[EAN code for a product]
4.
Mass
[nominal mass/single product mass]
5.
Tare
[tare value, set automatically when the product is selected out of the database]
6.
Min
[low limit for product weighing in CHECKWEIGHING mode]
7.
Max
[high limit for product weighing in CHECKWEIGHING mode]
8.
Tolerance [% value calculated in relation to mass, the parameter determines measuring area
for which the measurement is recognized as correct for DOSING mode.]