- 36 -
-
balance software detects necessity of carry out adjustment and signalizes it
by Celsius degree pictogram or by black colour pictogram on balance
display,
-
since pictogram occurrence, time interval of 5 minutes is counted within
which weighing procedure can be completed,
-
as the 5 minutes time elapses, balance display indicates message
CAL_30
and starts counting down from 30..29..28 to 0 (indicated value is the
counter),
-
balance user has 30 seconds to make a decision:
in order to start adjustment, do not take any actions,
in order to complete weighing procedure, press
ESC/TARE
key, If
pressed, balance returns to weighing procedure and displays last
weighing result. In 5 minutes balance will again indicate
CAL_30
message
-
the adjustment process can be postponed for multiple times, but it needs to
be stressed that long postponing of adjustment may be a reason of bigger
errors of weighing process. The errors are effect of temperature changes
and as a consequence changes of balance sensitivity.
Automatic adjustment process takes place in three different cases:
•
adjustment on plugging the balance to mains. After approximately 6
minutes from plugging the balance to mains, it automatically initiates
adjustment process,
•
adjustment triggered by temperature changes. Balance APPC series
is equipped with very precise system for monitoring temperature. At
each adjustment process, the temperature is saved in system. The
next adjustment is automatically initiated if temperature changes
more than 3
o
C from last saved temperature.
Summary of Contents for BALANCES APP/2C SERIES
Page 2: ...2 NOVEMBER 2012...
Page 22: ...22...
Page 70: ...70...
Page 71: ...71 Return to weighing mode...