- 81 -
16.2.
Printing date with specified date and time
Every weighing result can be printed with specification of measurement date
and time. The presence of date and time on a printout is conditioned by
connection of printers type
Kafka 1/Z
or
Kafka SQ S
to a balance.
When connected, go to balance menu and set parameters from group
P2 GLP
:
PdAt
-
YES
Ptin
-
YES
16.3.
Cooperation with statistic printer Kafka SQS
On connecting a statistic printer
KAFKA SQ S
to a balance, it is possible to
carry out statistics from completed measurements. An instance of a printout
including statistics from series of measurement:
1 9:02:15
+ 7.0016
g
2 9:02:39
+ 5.0152
g
3 9:02:58
+
12.0171
g
4 9:03:15
+ 9.9937
g
5 9:03:34
+ 12.0169
g
6 9:03:48
+ 22.0111
g
Date 13.09.2001 Time 9:04
n
6
number of samples
sum x
68.0556
g
total mass of samples
x
11.34260
g
average value
s
5.92328
g
standard deviation
srel
52.22
%
variance factor
min
5.0152
g
min value
max
22.0111
g
max value
R
16.9959
g
max – min difference
16.4.
Format of data sending
Weighing result can be sent to a peripheral device by pressing
key on a
balance keyboard or by setting a control command from computer level.
Summary of Contents for BALANCES APP/2C SERIES
Page 2: ...2 NOVEMBER 2012...
Page 22: ...22...
Page 70: ...70...
Page 71: ...71 Return to weighing mode...