- 150 -
carrying out, no parameter
available or incorrect format
LDS_I <CR><LF>
- command understood but not
accessible at this moment
n
– parameter, decimal value determining last digit settings
n
→
1 – always
2 – never
3 – when stable
Caution:
The numbering is assigned to a particular last digit option and it is identical for
all balance types.
The command changes settings for a current working mode if, for a particular
balance type, last digit settings are assigned to the working mode.
Example:
Command:
LDS_1 <CR><LF> – set last digit option to value ‘always’
Response:
LDS_OK <CR><LF> – ‘always’ value has been set
27.3.
Manual printout / Automatic printout
Moisture analyzer MA 3Y series enables generating manual or automatic
printouts.
•
Manual printout: press
key on stabilization of indication
(measurement result),
•
Automatic printout is generated automatically, in accordance to the settings
of automatic printout (
see point 11.6 of this user manual
).
The content of a printout depends on setting in menu <Standard printout> -
<Weighing printout template> (
see point 16.5 of this user manual)
.
Format of mass printout:
1
2
3
4 -12
13
14 15 16 17 18
stability
marker
space character mass space
unit
CR LF
Stability marker
[space] if stable
[?] if unstable
[!] air buoyancy correction mode is enabled
[^] if error of exceeding range to + occurs
[v] if error of exceeding range to – occurs