16
Determine start mass
-
-
Start mass determination (refer to
section 9.4.1).
Determine adjustment
factor
-
-
Adjustment factor determination (read
section 9.4.2).
Set defaults
-
-
Restoring default adjustment parameter
settings.
Linearity
-
-
Section linearity correction (refer to
section 9.5).
Weighing
-
-
-
Autozero range
0.25
0.1 ÷ 5
Autozero range (in divisions).
Stability range
1
0.1 ÷ 5
Stability range in [d] units.
Stability time
0.8
0.1 ÷ 5
Stability time, given in seconds.
Start mass control
1
0, 1, 2
Start mass control:
0
– disabled,
1
– range: from -5% to
+15% of start mass,
2
– range: from -
20% to +20% of start mass.
Information
-
-
-
A/D divisions
-
-
Current quantity of converter divisions.
Adjustment factor
100
-
Adjustment/calibration factor.
Start mass
100
-
Start mass value (in divisions).
Gcor
1
-
Gravitational correction factor (refer to
section 9.6).
Serial number
-
-
Serial number of the mass converter.
Adjustment qty counter
-
-
The quantity of factory parameter
recordings.
Factory start mass
100
-
Start
mass
value
expressed
in
converter's divisions, read in the course
of the factory adjustment.
Factory adjustment factor
100
-
Adjustment factor value, determined
during the factory adjustment.
Converter
-
-
-
Converter rate
6,25
6.25 ÷ 1600 Converter rate change.
9.3. Parameter Value Modification
Select a parameter, enter/modify its value.
Press
<Save>
button, message
<Save parameters?>
is displayed.
Press
<Yes>
button to confirm, message
<Changes saved>
is displayed.
Press
<OK>
button to confirm.