9.5. Linearity Correction 

 

„Linearity”

 tab enables access to thresholds and linearity deviations. 

 

 

 

Figure 9. Linearity correction window 

 

Where: 

 

 

Press  to  zero  all  threshold  values.  To  zero  the  deviations  and  the 
thresholds, select 

<Zero thresholds while zeroing linearity>

 option. 

 

Press to zero a value of a particular threshold/deviation. 

 

Press to determine deviation of a particular threshold. 

 

Procedure: 

 

  Enter linearity threshold values. 

  Press particular threshold button, 

, message 

<Load xxx>

 (where: xxx 

– entered mass value) is displayed. 

  Load the weighing pan with weight of the declared mass value and press 

<OK>

 button. 

  Deviation is calculated automatically and entered in the linearity threshold 

deviation field. 

  Determination  of  deviations  for  remaining  linearity  thresholds  proceeds 

analogously. 

Summary of Contents for MW-01-A

Page 1: ......

Page 2: ...2 JANUARY 2022...

Page 3: ...l reactions The device must not be operated in hazardous areas endangered with explosion of gases and in dusty environments In the case of damage immediately unplug the device from the mains The devic...

Page 4: ...cting Load Cell s Cable Shield 13 9 FACTORY PARAMETERS 14 9 1 Factory Parameter Access 14 9 2 Factory Parameters 15 9 3 Parameter Value Modification 16 9 4 Factory Adjustment 17 9 4 1 Start Mass Deter...

Page 5: ...nty does not cover mechanical damage caused by product exploitation other than intended damage of thermal and chemical origin damage caused by lightning overvoltage in the power network or other rando...

Page 6: ...The user is NOT ALLOWED to carry out any kind of repair of the device himself herself Any attempt of module modification repair etc by unauthorized persons will result with loss of validity of manufa...

Page 7: ...7 6 1 General View Figure 1 MW 01 A mass converter general view 6 2 Dimensions Figure 2 MW 01 A mass converter dimensions 6 3 Connectors Figure 3 Arrangement of connectors...

Page 8: ...r RS232 Pin1 NC Pin2 RxD Pin3 TxD Pin4 NC Pin5 GND Pin6 5VDC Pin7 NC Pin8 NC 6 5 Inputs Outputs Standard MW 01 A mass converter is equipped with 3 optoisolated inputs and 3 semiconductor outputs semic...

Page 9: ...tput type Solid state relay Maximum output current 0 5 A DC Maximum output voltage 30 VDC AC Input parameters Input quantity 3 Input type Optoisolated Input voltage range 5 24 VDC 6 6 Main Board Socke...

Page 10: ...cation unit 0 5 V Minimum load cell impedance 50 Maximum load cell impedance 1200 Load cell excitation voltage 5 VDC Load cell connection 4 or 6 wires SHIELD Platform quantity 1 Multi range option Yes...

Page 11: ...on a flat and even surface Keep it away from any sources of heat C Level the device Levelling procedure is to be found in TWM1 Scale user manual 7 2 START UP Plug the power supply to the mains Wait f...

Page 12: ...l signal REMARKS E SHIELD Refer to section 8 3 REF SENSE JP2 not to be soldered REF SENSE JP4 not to be soldered IN OUTPUT IN OUTPUT EXC INPUT EXC INPUT 8 2 4 Wire Load Cell Connection Connect 4 wire...

Page 13: ...erter in a metal housing connected with a platform via a load cell signal cable Scale of compact mechanical design mass converter in a metal housing connected with a platform via an extension arm etc...

Page 14: ...r Access Plug the power supply of the MW 01 A mass converter to the mains Press internal micro switch SW1 Figure 7 Factory parameters access switch Wait for connection to be established between the ma...

Page 15: ...001 0 001 50 Rounding of the last digit and position of a decimal point for weighing range I Reading unit range 2 0 001 0 001 50 Rounding of the last digit and position of a decimal point for weighing...

Page 16: ...quantity of converter divisions Adjustment factor 100 Adjustment calibration factor Start mass 100 Start mass value in divisions Gcor 1 Gravitational correction factor refer to section 9 6 Serial num...

Page 17: ...o converter s memory 9 4 2 Adjustment Factor Determination Unload the weighing pan Press Determine adjustment factor button message Unload the pan is displayed Press OK button to confirm start mass de...

Page 18: ...lue of a particular threshold deviation Press to determine deviation of a particular threshold Procedure Enter linearity threshold values Press particular threshold button message Load xxx where xxx e...

Page 19: ...scale is calibrated at the place of use then the gravitational correction value g cor parameter is 1 00000 If the scale is calibrated away from the place of use longitudinal change the value of g cor...

Page 20: ...the 309R module an extra M12 8P connector is installed on the housing Figure 10 RS485 module 10 2 1 RS485 Pins RS485 Pin1 B Pin2 NC Pin3 A Pin4 A Pin5 NC Pin6 NC Pin7 B Pin8 NC 10 3 ETHERNET Module T...

Page 21: ...dule combined with AB6003 module expands the functionality of the MW 01 A converter by Ethernet IP interface The module is intended to be mounted inside the indicator For the 494R module a hermetic RJ...

Page 22: ...13 495R current loop module The module s working mode is set using S1 switch when setting the module s working mode refer to the below table R1 R2 WORKING MODE 0 0 0 10 V 1 0 4 20 mA 0 1 0 20 mA 1 1 0...

Page 23: ...OLOURS Colour Signal white IOUT grey GND yellow 24V DC green GND 10 5 2 Technical Specifications Output resolution 16bit Linearity error 0 01 4 mA indication error 0 1 Temperature error for 4 mA 25 pp...

Page 24: ...ing with coding respective for the PROFIBUS standard Figure 14 PROFIBUS module 10 6 1 PROFIBUS Pin Arrangement PROFIBUS IN male Pin1 NC Pin2 A Pin3 NC Pin4 B Pin5 NC PROFIBUS OUT female Pin1 5V Pin2 A...

Page 25: ...25 Figure 15 Profinet Module 11 ACCESSORIES MW 01 A computer cable PT0020 MW 01 A Ethernet cable PT0212 MW 01 A 3IN cable PT0256 MW 01 A 3OUT cable PT0256 MW 01 A RS485 cable PT0383...

Page 26: ...26...

Reviews: