22
10.5. Analog Output Module
The 495 R module expands the functionality of the MW-01-A converter by an
analog output, 4-20 mA, and a voltage output, 0-10 V. The module is intended
to be mounted inside the MW-01-A. The 495R module is a passive module.
The MW-01-A converter's housing features a cable gland through which a
3-metre cable 6x0.34 mm2 is fed, which cable is terminated with isolated wires.
Figure 13. 495R current loop module
The module’s working mode is set using
S1
switch, when setting the module's
working mode refer to the below table.
R1
R2
WORKING MODE
0
0
0-10 V
1
0
4-20 mA
0
1
0-20 mA
1
1
0-24 mA
Analog output module board provides description of settings, for the description
look around
S1
switch.
BY DEFAULT, SET THE ANALOG OUTPUT MODULE TO 4-20 mA MODE.
Analog output must be calibrated (adjusted); using P1 potentiometer set the
correct offset (e.g., for 4 - 20 mA output, adjust current so that for 0 kg
indication it is precisely 4 mA). In the case of remaining operation modes
proceed analogously.