15
Figure 8. Factory parameters window
9.2. Factory Parameters
NAME
DEFAULT
VALUES
REMARKS
Adjustment
-
-
-
Range
3.009
-
Maximum range + 9e overload.
Range 2 threshold
0
-
Switching point of range II. For single
range balances set
0
.
Range 3 threshold
0
-
Switching point of range III. For I-range
or II-range scales set
0
value.
Reading unit: range 1
0.001
0,001
50
Rounding of the last digit and position
of a decimal point for weighing range I.
Reading unit: range 2
0.001
0,001
50
Rounding of the last digit and position
of a decimal point for weighing range II.
Reading unit: range 3
0.001
0,001
50
Rounding of the last digit and position
of a decimal point for weighing range III.
Adjustment weight
3
-
Adjustment weight mass.
Adjustment unit
kg
g, kg, lb
Adjustment (calibration) unit.
Verified
NO
YES - NO
Declaring whether the scale is verified
or not.
YES
- verified scale,
NO
- non-
verified scale.