17
9.4. Factory Adjustment
9.4.1. Start Mass Determination
Unload the weighing pan.
Go to
<Adjustment>
tab, press
<Determine start mass>
button,
message
<Unload the pan>
is displayed.
Press
<OK>
button to confirm, start mass adjustment begins.
Upon completion, message:
<Start mass determination completed
successfully>
is displayed.
Press
<OK>
button to confirm.
Press
<Save>
button, message:
<Save parameters?>
is displayed.
Press
<Yes>
button to confirm, message
<Changes saved>
is displayed.
Press
<Yes>
button to confirm. Determined start mass gets saved to
converter's memory.
9.4.2. Adjustment Factor Determination
Unload the weighing pan.
Press
<Determine adjustment factor>
button, message:
<Unload the
pan>
is displayed.
Press
<OK>
button to confirm, start mass determination begins.
Upon completion, message:
<Load xxx>
(where: xxx
– declared
adjustment weight mass) is displayed.
Load the weighing pan with weight of the specified mass value and press
<OK>
button.
Upon completion, message:
<Adjustment factor determination
completed successfully>
is displayed.
Press
<OK>
button to confirm.
Press
<Save>
button, message:
<Save parameters?>
is displayed.
Press
<Yes>
button to confirm, message
<Changes saved>
is displayed.
Press
<Yes>
button to confirm. Determined adjustment factor gets saved
to mass converter's memory.