19
Press
<Save>
button to save changes, message
<Save parameters?>
is displayed.
Press
<Yes>
button to confirm, message
<Changes saved>
is displayed.
Press
<Yes>
button to confirm.
9.6. Gcor
The function of gravitational correction compensates changes of earth gravity
being a result of different latitude. It enables correct calibration of the weighing
device away from the place of use. The gravitational correction value must be
entered with reference to tables prepared by „Radwag Wagi Elektroniczne”
or calculated using the below formula:
kal.
uzy t.
g
g
Gcor
Correction value ranges between 0.90000 ÷ 1.99999.
If the scale is calibrated at the place of use, then
the gravitational correction value (<g-cor.> parameter) is
1.00000. If the scale is calibrated away from the place of use
(longitudinal change), the value of <g-cor.> parameter must
be corrected.
10. OPTIONAL EXTENSION MODULES
The MW-01-A mass converter can be equipped with additional optional
interface modules: ETHERNET, RS485, PROFIBUS, PROFINET, ETHERNET
IP and analog outputs (4-20 mA or 0-10 V). The modules are installed inside
the device. Optional interfaces are installed interchangeably.
10.1. Exclusions Regarding Installation of Additional Modules
INSTALLED MODULE
EXCLUDED MODULES
ETHERNET
RS485, PROFIBUS, ETHERNET IP, PROFINET
PROFIBUS
RS485, ETHERNET, ETHERNET IP, PROFINET
PROFINET
RS485, PROFIBUS, ETHERNET IP, ETHERNET
RS485
ETHERNET, PROFIBUS, ETHERNET IP, PROFINET