8. BASIC OPERATIONS
8.1. Adjustment (for models whit external adjustment)
The adjustment mass standard does not come with the balance.
Prior to adjustment prepare the right mass standard. You are
recommended to use a mass standard of class F1 or F2. View
the table and select the mass standard you need.
External adjustment is carried out using an external mass standard of the right
accuracy and weight value, which value depends on balance type and capacity.
Balance
model
Mass
standard [g]
PM 15.4Y
10000 g
PM 60.1.4Y
50000 g
PM 60.05.4Y
50000 g
PS 3000.3Y
5000 g
PS 10100.3Y.M
5000 g
WLY 1/D2
1000 g
WLY 2/D2
2000 g
WLY 10/D2
10000 g
WLY 12/F1
10000 g
WLY 20/D2
20000 g
WLY 30/F1
20000 g
WLY 120/C2
100000 g
1
3
2
4
Press “Parameters” button.
Press “External Adjustment” parameter.
Press “Adjustment” button.
Unload the weighing pan and press
button for confirmation.
0%
Home
0.0000
100%
Admin
Start drying [Accept]
Weighing
Supplementary unit
Tare
Gross
0.000 g
0.000 g
Packaging
1
2
User adjustment
Adjustment weights
3
Adjustment
External adjustment
External adjustment
Parameters
Database
Communication Peripherals
Adjustment
Operators
Profiles
Adjustment
Unload the weighing pan
Adjustment