- 25 -
automatically and its result is shown on the display (if the balance is connected via
communication port with a computer or a printer, the adjustement result will be printed out).
Press ESC to go back to the previous menu.
•
P1.5 AUTO. CALIB. T. – Automatic adjustment
Determination of factor which signals start of automatic internal adjustment (function
unavailable for verified balances).
NONE
neither time nor temperature is to trigger adjustment (adjustment
automatically turned off)
TEMP
adjustment triggered by change of temperature
TIME
adjustment triggered by time interval set in P1.6 AUTO CALIB. C.
BOTH
adjustment triggered by time or temperature.
•
P1.6 AUTO CALIB. C. - Time of automatic adjustment
Determination of time interval for innitiation of automatic adjustment process (function
unavailable for verified balances).
7.3. MANUAL ADJUSTMENT
7.3.1. Internal Adjustment
The balance performs internal adjustment automatically. During adjustment process do not load the
weighing pan with any load. The message <
DO NOT TURN OFF CALIBRATION
> is displayed in the
bottom line. Once adjustment process is completed, balance saves its result in memory and returns
to weighing mode.
CAUTION!
- Press Esc button to abort adjustment process.
- If a weighing pan is loaded, message informing about an error turns out on the display. Adjustment
process is aborted, balance returns to the weighing mode. Adjustment process may be repeated upon removal
of the excess load.
7.3.2. External Adjustment
The external adjustment for PS series balances should be carried out with an external mass standard
of class F
1
(function unavailable for verified balances).
Procedure:
•
Run an external adjustment process, the balance displays a command ordering to unload the
weighing pan,
<REMOVE MASS>
(the weighing pan must be empty). When the weighing pan
is unloaded, press
button.
•
The balance determines mass of an empty pan, message <
CALIBRATION
> is displayed in the
bottom line. Next, message <
PLACE MASS
> and mass value to be placed on the weighing
pan are displayed,
e.g.
200.000g
(depending on the type of balance).
•
Place an external adjustment weight of displayed mass value and press
button
.
The
balance determines the mass, message <
CALIBRATION
> is displayed in the bottom line. On
completing adjustment process the balance returns to submenu
P1.2 EXT.CALIB.
7.3.3. User Adjustment
The external adjustment for PS series balances should be carried out with an external mass standard
of class F
1
(function unavailable for verified balances).
Procedure:
•
Run an external adjustment process, the first step of the process is to declare the mass of a
weight that is to be used for adjustment. The mass must be
≥
30% Max capacity.