- 53 -
11.7.1. Additional Settings
Apart from standard settings for this mode (i.e. weighing mode settings) some additional parameters
have been designed to describe operation of the mode.
Additional settings list:
•
AVERAGING TIME
– Amount of time during which recorded measurement results are
analysed. Obtained data is used to determine average measurement result.
•
THRESHOLD
- It is a value expressed in mass measuring units. In order to start measurement,
the indication value has to exceed the set threshold value.
•
AUTOSTART
- Determines start criterion for measurements: whether they are to be initiated
manually, on pressing a button or selecting START, or automatically.
When the parameter is set to <YES> option, the measurement starts automatically if an
indication on a balance display exceeds the set threshold value. The following measurement
can start on removing the weighed object from a balance pan (the indication has to return
below the value set in the threshold), and on loading the pan with a new object at a moment
when the set threshold value indication is exceeded.
Prior to tests it is necessary to adjust settings of the above options so that correct values were
provided. The selected options shall be respective to needs and expectations resulting from working
environment.
F Shortcut Keys
You can declare which particular weighing function is to be launched using F1, F2, F3 or F4 keys.
For procedure informing you how to declare the particular function, read
F Shortcut Keys
section.
11.7.2. Running the Process Manually – Means of Operation
To start the process of weighing manually, enter the mode settings, select an <AUTOSTART>
parameter and set <NO> value.
Procedure:
After altering the settings, return to the main window by pressing
button repeatedly.
Next, set the averaging time in seconds. It is the time, during which the balance software acquires
measurements and based on these measurements the average result is determined.
The <THRESHOLD> parameter is not required for this measurement mode.
Select <ANIMAL WEIGHING> mode.
Place the container in which the measurement is to be carried out on the weighing pan and when the
indication is stable carry out taring.
Next, enter the mode options and start the measurement procedure following the instruction below.