- 86 -
Response options:
C1_I CR LF
- command understood but not accessible at this moment
C1_A CR LF
- command understood and in progress
MASS FRAME
- response: mass value in a basic measuring unit
Response format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S
I
space
stability
marker
space
character
mass
space
unit
CR
LF
Switch off continuous transmission in a basic measuring unit
Format:
C0 CR LF
Response options:
C0_I CR LF
- command understood but not accessible at this moment
C0_A CR LF
- command understood and carried out
Switch on continuous transmission in a current measuring unit
Format:
CU1 CR LF
Response options:
CU1_I CR LF
- command understood but not accessible at this moment
CU1_A CR LF
- command understood and in progress
MASS FRAME
- response: mass value in a current measuring unit
Response format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S
U
I
stability
marker
space
character
mass
space
unit
CR
LF
Switch off continuous transmission in a current measuring unit
Format:
CU0 CR LF
Response options:
CU0_I CR LF
- command understood but not accessible at this moment
CU0_A CR LF
- command understood and carried out
Set min checkweighing threshold
Format:
DH_XXXXX CR LF
, where:
_ -
space
,
XXXXX
– mass format
Response options:
DH_OK CR LF
- command carried out
ES CR LF
- command not recognised (mass format incorrect)
Set max checkweighing threshold
Format:
UH_XXXXX CR LF
, where:
_ -
space
,
XXXXX
- mass format
Response options:
UH_OK CR LF
- command carried out
ES CR LF
- command not recognised (mass format incorrect)
Give value of min checkweighing threshold
Format:
ODH CR LF
Response:
DH_MASA CR LF
- command carried out
Response format:
1
2
3
4-12
13
14
15
16
17
18
19
D
H
Space
mass
space
unit
space
CR
LF
Mass
- 9 characters with right justification
Unit
- 3 characters with left justification
Give value of max checkweighing threshold
Format:
OUH CR LF
Response:
UH_MASA CR LF
- command carried out
Summary of Contents for PS 200/2000 R2