- 91 -
Response options:
RV_A_”x” CR LF
- command understood, response: program version
RV_I CR LF
- command understood but not accessible at this moment
x
– program version (in between inverted commas)
An example:
Command:
RV CR LF
– return program version
Response:
RV_A_” 1.1.1”
– program version: ”1.1.1”
Set AUTOZERO function
Format: A_n CR LF
Response options:
A_OK CR LF
– command carried out
A_E CR LF
– error in-course of command execution, no parameter or incorrect format
A_I CR LF
– command understood but not accessible at this moment
n
– parameter, decimal value determining autozero settings
n →
0 – autozero off
1 – autozero on
CAUTION!
Command changes settings for a current working mode.
An example:
Command:
A_1 CR LF
– turn autozero function on
Response:
A_OK CR LF
– autozero function is on
AUTOZERO function operates until it is turned off by A 0 command.
Set ambient conditions state
Format: EV_n CR LF
Response options:
EV_OK CR LF
– command carried out
EV_E CR LF
– error in-course of command execution, no parameter or incorrect format
EV_I CR LF
– command understood but not accessible at this moment
n
– parameter, decimal value determining ambient conditions state
n
→
0 – unstable ambient conditions
1 – stable ambient conditions
CAUTION!
Command changes settings for a current working mode.
An example:
Command:
EV_1 CR LF
– set value ‘stable’ for ambient conditions option
Response:
EV_OK CR LF
– ambient conditions option set to value ‘stable’
<AMBIENT CONDITIONS> parameter is set to value <STABLE> until command EV 0 swaps it to
value <UNSTABLE>.
Set filter
Format: FIS_n CR LF
Response options:
FIS_OK CR LF
– command carried out
FIS_E CR LF
– error in-course of command execution, no parameter or incorrect format
FIS_I CR LF
– command understood but not accessible at this moment
n
– parameter, decimal value determining filter number
n →
1 – very fast
2 – fast
3 – average
4 – slow
5 – very slow
CAUTION!
The numbering is assigned to a particular filter name and it is identical for all balance types.
The command changes settings for a current working mode if, for a particular balance type, filter settings are
assigned to the working mode.