- 59 -
9.
Next, take the sinker out of the hook, put the beaker with liquid on the beaker’s basis (the
beaker cannot touch the hook), gently hang the sinker on the hook providing that it is totally
immersed in the tested liquid (sample mass determination in liquid) – and confirm the indication
when stable
.
10. On the second confirmation of the indication, the program automatically calculates and displays
the tested solid object density. At the same time, a report on the measurement is sent to a
selected port of a printer.
An exemplary report:
The report may be reprinted upon pressing
button. In order to finish, press
button. The
software returns to the main window of the mode and a next measurement can be performed. The
balance operates with previously set parameter values (liquid, temperature) thus shortening time
needed for the exact measurement.
11.10. STATISTICS
Statistics is a working mode allowing to acquire data from series of measurements and to produce
statistics using the acquired data. Settings of this function determine which data is displayed.
Summary of Contents for PS 200/2000 R2
Page 95: ...95...