- 147 -
Once you select the mode, the function is active and information about
the net mass is displayed in a bottom line (only if a different information
has not been selected by a user).
In order to provide proper operation, set the threshold in grams
determining point beyond which the function starts to register
maximum force apllied.
•
From now on the balance registers and holds every single weighment
which is above the threshold, and which is higher than the result of the
previous peak hold. If the software detects mass above the threshold,
the highest detected indication is held on the main display and the
pictogram <Max> is shown at the top of the display.
A user can print the result out by pressing
button.
The start of the next process of peak hold measurement is possible
only after removing the load from the weighing pan and pressing
button. This causes returning to the main window of <PEAK
HOLD> mode, pictogram <Max> is automatically deleted.