- 178 -
Enable automatic internal calibration of the balance
Format:
IC0 CR LF
Accessible responses:
IC0_I CR LF
- command understood but not accessible at this moment
IC0_OK CR LF - command carried out
Operation is disabled for verified balances.
Lock balance keypad
Format:
K1 CR LF
Accessible responses:
K1_I CR LF
- command understood but not accessible at this
moment
K1_OK CR LF
- command carried out
Command locks the balance keypad (proximity sensors, touch panel) until the
moment of turning the balance off or until sending K0 command.
Unlock balance keypad
Format:
K0 CR LF
Accessible responses:
K0_I CR LF
- command understood but not accessible at this
moment
K0_OK CR LF
- command carried out
OMI - Give available working modes
Command overview:
Command returns accessible working modes.
Format:
OMI <CR><LF>
Accessible responses:
OMI <CR><LF>
n_”Mode name” <CR><LF>:
n_”Modename” <CR><LF>
OK <CR><LF>
- command carried out, accessible working
modes returned
OMI_I <CR><LF>
- command understood but not accessible at
this moment