- 30 -
9.4.3.
Autozero function
The software features autozero function (
Auto
) ensuring precise mass
indication. This function automatically controls and corrects zero indication.
When Autozero is enabled, it compares balance indications at declared time
interval e.g. 1s, providing that weighing pan is unloaded and display indication
is close to zero. If results vary less than declared AUTOZERO range e.g. one
division, balance zeros automatically, marker of stable measurement result –
and precise zero marker –
are displayed
.
If AUTOZERO function is enabled, then each weighing process starts from
precise zero point. There are, however, some cases when this function can be a
disturbing factor for the measuring process; e.g. very slow placing of load on the
weighing pan (load pouring). Here, zero indication correction can also correct
actual indication of loaded mass.
9.4.1. Displaying the last digit
Function enables displaying the last digit on a balance display.