- 31 -
9.4.2.
Balance ambient conditions
This parameter has two settings: STABLE and UNSTABLE. Selecting STABLE
mode makes the balance work much faster, ie. weighing takes much less time
than for UNSBABLE mode. This parameter relates to ambient and
environmental conditions in which the balance operates. If the ambient
conditions are unstable it is recommended to use UNSTABLE mode. Normally
the parameter is set to STABLE option.
9.5. AUTOTARE
In case of loads with different tare values being weighed one after another,
automatic tare function is useful for quick determination of the weighed loads
net mass.
When function is active (<P2.2.2 AUTOTARE> value set to <YES>), the
operating process is as follows:
•
Make sure that the weighing pan is empty and press button
responsible for zeroing,
•
Put product packaging on a weighing pan,
•
After measurement stabilization
automatic tarring
of the packaging
mass will proceed (
Net
marker appears in the upper part of the
display),
•
Put product that is to be packed on the packaging;
•
Display will show a net mass of the product;
•
Take off the product together with the packaging;
•
The balance cancels tare value (packaging weight saved in balance
storage during the first step of the operating process) after the gross
mass value (set in
< P 2.2.3.2 AUTO THRES>
parameter) has been
exceeded;
•
Put packaging of a next product on a weighing pan, after measurement
stabilization automatic tarring of the packaging weight proceeds (
Net
marker appears in the upper part of the display);
•
Put a next product that is to be packed.