- 55 -
The first step of the process is to declare the mass of a weight that is to be used
for adjustment. The mass must be
≥
30% Max capacity.
Once the mass of the weight is entered and confirmed, the message prompting
the user to remove the weight from the pan is displayed:
<REMOVE MASS>
(the weighing pan must be empty). After following the instruction press
button, the balance determines the weight of an unloaded pan, message
<
CALIBRATION
> is shown in the bottom line. Next, message <
PLACE MASS
>
and mass value to be placed on the weighing pan are displayed,
e.g.
200.000g
(depending on the type of balance). Place an external adjustment weight of
displayed mass value and press
button
.
The balance determines the
mass, message <
CALIBRATION
> is displayed in the bottom line. On
completing adjustment process the balance returns to submenu
P1.2
EXT.CALIB.
If a balance is verified, it is not possible for a user to carry out the
external adjustment process.
13.4. Adjustment report printout
At the end of each adjustment process or adjustment test, an adjustment report
is generated automatically and sent to communication port COM 1. The content
of the report is declared in menu P5.1 CAL REPORT.
The instruction on how to declare settings for this option is described in section
about printouts.
The report can be printed out via a printer connected to the balance or it can be
sent to the computer and saved as a file for archiving purposes.