69
15.4.1. Format of sent data after pressing the PRINT button
Depending on setting of
P4.5 PStb
parameter only stable measurement or
mass of instantaneous mass will be sent.
For verified balance printout of temporary measurements will be blocked
for instantaneous measurements.
Format of printout
1
2
3
4 - 12
13
14 - 16
17
18
stability
marker
space
mark
mass
space
unit
CR
LF
stability marker – [space] if stable
[?] if unstable
[^] if there is an error of exceeding the range on +
[v] if there is an error of exceeding the range on -
Mark
– [space] for positive values or [-] for negative values
mass
– 9 marks alignment to the right
units
– 3 marks alignment to the left
15.4.2. Format of sent data for orders generated from the computer
Balance after receiving the command answers first:
XX_A CR LF – commend understood, realization started
XX_I CR LF
– commend understood, but unavailable at the moment
XX _ ^ CR LF
– commend understood, but max range is exceeded
XX _ v CR LF
– commend understood, but min range is exceeded
XX _ E CR LF
– error occurred while command realization – exceeded time limit
while waiting for stable result (time limit is characteristic
parameter of balance)
XX
– name of command
And afterwards
Summary of Contents for PS Series
Page 2: ...2 JUNE 2008...
Page 35: ...35...
Page 48: ...48 if you want to choose one them e g 20pcs press ENTER PRINT and proceed as shown on picture...
Page 58: ...58...