background image

 

 

1.   GENERAL INFORMATION 

 
 

THB module is a measuring device cooperating with laboratory balances of 3Y, 4Y, X2 

series, PCs and ECR recorders. The  THB module enables measurement of ambient 
conditions parameters (temperature, humidity and atmospheric pressure) at the workstation. 

The indications may be displayed either on an indicator screen or a computer screen. 
On the basis of measurement  values transferred from the THB module, the 3Y, 4Y, X2 

balance program determines air density.  The determined air density is displayed on the 
screen too. 

THB.PC module (PC version) is operated using „THB-R Single” software. 
THB.R module (ECR recorder version) is operated using „THB-R Multi” software. 

 

2.  THB MODULE DESIGN 

 
THB.x module: 

 
1 – housing, 
2 – sensors shield, 
3 – connector for an additional temperature sensor (TH-W2), 
4  –  cable terminated with a respective connector  (connector type conditioned by THB 
module model): 
      DSUB15 – for THB.Y; 

DB9 – for THB.X and THB.PC; 
PS2 – for THB.R. 
 

 

 

 

Summary of Contents for THB.Y

Page 1: ...THB Ambient Conditions Module IMMU 49 01 10 17 EN USER MANUAL THB Y THB X THB PC THB R www radwag com...

Page 2: ...2 OCTOBER 2017...

Page 3: ...2 THB MODULE DESIGN 4 3 UNPACKING 5 3 1 Unpacking 5 4 START UP 6 4 1 THB Y and THB X Modules 6 4 2 THB PC Modules 8 4 3 ECR Recorder and THB R Set 9 5 MAINTENANCE SERVICE RECYCLING 10 5 1 Maintenance...

Page 4: ...f measurement values transferred from the THB module the 3Y 4Y X2 balance program determines air density The determined air density is displayed on the screen too THB PC module PC version is operated...

Page 5: ...B B connector recorder computer connection via standard USB cable d PS2 connector for connection of liquid aspirator weighing table for pipettes calibration e 2 x PS2 connector for connection of THB R...

Page 6: ...et communication parameters THB Y THB Y has been designed to enable measurement of ambient conditions It cooperates with laboratory balances of 3Y and 4Y series It is connected directly to COM 2 port...

Page 7: ...It cooperates with laboratory balances of X2 series It is connected directly to either COM 1 or COM 2 port of the balance THB X allows to measure air temperature atmospheric pressure air humidity Set...

Page 8: ...ns provided in user manual of the balance THB PC THB PC has been designed to enable measurement of ambient conditions It is connected directly to a computer port via RS232 Additionally it is necessary...

Page 9: ...eadout of metrological data simultaneously from all the connected modules within a specified time interval The data is saved to the recorder s internal memory The recorder is powered by an external 10...

Page 10: ...1 Maintenance Power supply housing and sensors shield can be cleaned using cloth soaked with alcohol based substance 5 2 Service Only authorized RADWAG service points are allowed to disassembly repair...