After selecting the required value save changes according to 29.3.2
of the manual.
29.6. Autozero function
The autozero function has been introduced in order to assure precise
scale’s indications. This function controls and corrects
„0”
indication.
While the function is active it compares the results successively with
constant frequency. If results differ less than the declared value of autozero
range, for example 1unit, the scale will be automatically zeroed and the
stability result markers
and zero indicator
will be displayed.
When AUTOZERO function is active, each measurement starts with
precise zero. In particular cases the function can disturb measuring
process. For example when product is loaded very slowly on scale
pan (pouring),
Autozero
system can influence real weighing results.
Procedure:
•
Readout parameters value according to 29.3.1 of the manual,
•
Press, then rewind
Autozero
parameter window.
- 105 -
Summary of Contents for TMX series
Page 2: ... 2 SEPTEMBER 2011 ...