24.8.1. Local settings
Local settings for
<
Differential weighing>
function are to be accessed
in submenu:
„
Working modes /
Weighing /
Differential
weighing
:
Activation
Activation of differential weighing function
(
- function active,
- function inactive)
Batch portion
type
Batch type, for differential weighing:
Value
– differential weighing is realized according to
declared as measuring series batch value.
Filter
–
differential weighing is realized according to filter and
batch value declared as a measuring series value.
Filtering
Filter type declaration, differential weighing is realized
according to this filter.
Values:
Product, Client, lot
number, batch number, source warehouse, target
warehouse, packaging. Parameter
<Filtering>
is not
accessed if
<Batch portion type>
parameter value is
set to
<Value>
.
Charge
Measuring series value for differential weighing
24.8.2. Reporting realized processes of differential weighing
Report of each realized process of differential weighing is saved to database
<
Differential weighments reports>
, date and hour of an executed
process is the file name. To view list of data for a process of differential
weighing see point 35.5.8 of this manual.
24.9. Peak hold
The scale enables measuring the maximum force influencing the weighing
pan.
Procedure:
•
Enter
<
Working modes>
parameter group as described
in chapter 24 of this manual,
116