26.2.1. Automatic correction of reference mass
It concerns the <
Parts counting> working mode
Working mode
<
Parts counting>
comprises a special function
<
Automatic correction of reference mass>
, that can be used for
correcting the unit mass
<SMP>
. To enable the function in parameters you
need to:
•
Enter parameter group
<
Working modes>
according to ch.
24 of this manual, choose:
„
Parts counting /
Automatic
correction of reference mass”
and set appropriate option.
Options:
- Automatic correction of reference mass disabled
- Automatic correction of reference mass enabled
Function
<
Automatic correction of reference mass>
in mode
<
Parts counting >
is enabled at the moment of estimating the sample
quantity and signalled by displaying
<PCS>
and
<SMP>
(single part/piece
mass) on the top part of the display.
There are four criteria of working “Automatic correction of reference
mass”
function:
1. equilibrium should be reached,
2. quantity of parts/pieces should be increased,
3. added quantity of parts/pieces should not be greater than double
number of parts/pieces on the pan,
4. the new sample can be different from the old sample by
±
0.3
of pcs (absolute value),
If a user recognises that the sample quantity is adequate the unit mass
(single piece mass) can be saved (see ch. 26.6 of this manual)
and disable the function by pressing
.
121