Report number pattern:
X / y y / M M / d d / H H / m m / s s / T
, where:
X - Test mode:
U – regulation compliant test,
Z – test terminated by an operator,
yy – the last two digits of year,
MM - month,
dd - day,
HH - hour,
mm - minute of test completion,
ss - second of test completion,
T – average tare estimation.
Users can view data from subsequent average tare tests
.
Procedure:
•
Enter
submenu
<
Reports>
according to ch. 35 of this manual,
•
Enter
database
<
Average Tare Reports>
and choose a required
one.
Inventory of fields in the average tare database:
Product
Product name which packages are tested
Status
Test Status/Result
Date
Date of performing procedure
Tare
The result
S
Standard deviation
0.25 T1
Condition to fulfill
Number of
measurements
ents of subsequent
Number of measurem
packages
Operator
An operator name
231