247
Example:
S I CR LF
– computer command
S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF
- command done, mass value
in basic unit is returned immediately.
37.4.7. Send immediate results from all platforms in basic units
Syntax:
SIA CR LF
Possible answers:
SIA_I CR LF
- command comprehended but cannot be executed at the moment
MASS FRAME „P1” CR LF
MASS FRAME „P2” CR LF
- mass values are immediately returned from all
platforms in basic units
Frame format with mass from subsequent platforms as indicator reply:
1 2
3
4
5
6 7-15 16 17 18 19 20 21
P n space stability space sign mass space
unit
CR LF
n
- weighing platform number
mass
- 9 characters justified to the right
unit
- 3 characters justified to the left
Example:
Let us assume that both platforms are connected to scale.
S I A CR LF
– computer command
P 1 _ ? _ _ _ _ _ _ 1 1 8 . 5 _ g _ _ CR LF
P 2 _ _ _ _ _ _ _ _ _ 3 6 . 2 _ k g _ CR LF
- command done, mass values
from both platforms are returned in basic units
37.4.8. Send the stable result in current unit
Syntax:
SU CR LF
Possible answers: