18.2. Configuration of outputs
Ascribing a function to the output enables the output at the same time.
If an output has no ascribed function it is disabled.
Procedure:
•
Enter
<
Inputs / Outputs>
according to ch. 18 of this manual,
•
Choose
<
Outputs>
and enter the required output, then you
will see the list of functions:
None
Output disabled
Stabile
Stable weighing result over LO threshold value
MIN stable
Stable weighing result below the MIN threshold
MIN non-stable
Non-stable weighing result below the MIN threshold
OK stable
Stable weighing result between MIN and MAX
thresholds
OK non-stable
Non-stable weighing result between MIN and MAX
thresholds
MAX stable
Stable weighing result over the MAX threshold
MAX non-stable
Non-stable weighing result over the MAX threshold
Zero
Stable weighing result zero net
Confirmation of
cycle completion *
Signal that confirms that a cycle of dosing has been
completed (the defined amount)
Zero
Zero weighing result („zero” indication)
! OK unstable
Unstable weighing result out of OK threshold
! OK stable
Stable weighing result within OK threshold
Measurement
saved
Signal for measurement record confirmation -
OUT programmed to be active for 500 [ms]
Tarring completed
Signal for tarring operation confirmation -
OUT programmed to be active for 500 [ms]
Zeroing completed
Signal for zeroing operation confirmation -
OUT programmed to be active for 500 [ms]
Error
Error message displayed
PGC – pending
weighing process
Signal for awaiting for weighing performed within
PGC control process
84